- AC/DC Adapters[4]
- Testing Equipment[10]
- Other Electronic Measuring Instruments[10]
- Other Measuring & Gauging Tools[1]
- Other Measuring & Analysing Instruments[10]
- Power Accessories[1]
- Instrument Parts & Accessories[4]
- Magnetic Materials[2]
- Other Machinery[1]
- Other Machine Tools[1]
- Other Metal Processing Machinery[1]
- Other Electronic Components[2]
- Uninterrupted Power Supply (UPS)[1]
- Transformers[1]
- Contact Person : Mr. Rong Justin
- Company Name : Hunan Linkjoin Technology Co., Ltd.
- Tel : 0086---15197866786
- Fax : 0086-0738-8326398
- Address : Hunan,Loudi,5th floor, Carve Out Building, Economic Development Zone, Loudi
- Country/Region : China
- Zip : 417000
AC Hysteresisgraph - Model MATS-2010M
Detailed Product Description
Automatic measurement of the dynamic magnetic hysteresis loop of oriented and non-oriented silicon steel under the condition of 50 Hz, 60 Hz, 400 Hz and 1 kHz, accurate measurement of the static magnetic characteristic parameters such as amplitude permeability μa, loss angle δ, specific total loss Ps, remanence Br and coercivity Hc. With special testing fixture, it can directly measure the iron core and stator core of the transformer.
Windows measurement software applied simply. The product conforms to China National Standards GB/T 3655-2000, GB/T 13789-92 and international standard IEC 404-2, IEC 60404-3, IEC 60404-6.
The controlling of computer and A/D sampling replace the traditional analog bridge, frequency meter, ammeter, voltage meter, power meter, the whole testing process is automatic completed.
General Features
·Testing sample varieties: hot rolling, cold rolling silicon steel materials, permalloy, amorphous and nm crystal.
·Testing sample shapes: ribbon and chip open samples, annular, E and U closing samples.
·Open circuit samples adopt Epstein Square to form closed magnetic circuit, it can also select permeameter. Closing sample can be put under direct winding measurement, finished transformer can be also put under direct measurement.
·Sample (iron core), magnetizing coil (N1) and measuring coil (N2) form a no-load transformer.
·Non-inductive resistance connect to the magnetizing coil loop, and measure the pressure drop of non-inductive resistance to determine the magnetizing current and get the field intensity. The field peak value can be locked through the data feedback, the precision is 0.5%.
·To obtain the field induction intensity through the numeral integral in the voltage of measuring coil. The field peak value can be locked through the data feedback, the precision is 0.5%.
·Power source and sampling amplifier put into a chassis which make the interface very simple: a RS232 interface connect to the computer, the second voltage signal connect to high speed A/D card.
·Adopt volammetry and digital integration to measure dynamic hysteresis loop, it can accurately measure dynamic magnetic characteristic parameters such as μa, δ, Ps, Br and Hc, etc.
·Automatic and continuous measurement up to 255 testing points, the testing time of every testing point is about 30 seconds, fixed-frequency, fixed Bm or fixed Hm optional in multipoint test.
·Powerful function of the software, the technical request to the testing staff is very low.
System Specifications
Under 50 Hz frequency, use 25 cm Epstein Square to measure silicon steel sample, technical indices as follows:
Parameters measured | Bm (%) | Hm (%) | μa (%) | Ps (%) | δ (%) |
Uncertainty (k = 2) | 1 | 1 | 2 | 1 | 2 |
Repeatability (constant temperature) | ± 0.5 | ± 0.5 | ± 1 | ± 0.5 | ± 1 |
Instrument Specifications
MATS - 2010M Dynamic Hysteresisgraph | PC6112 A/D Card | |
Output Power: 500 VA sine wave Frequency Range: 45 Hz ~ 1000 Hz Frequency Fineness: 1 Hz Frequency Error: < 0.05% Output Voltage: 0 ~ 10 V ~ 50 V ~ 150 V ~ 300 V, four automatic ranges Voltage Fineness: Program Control 1 mV, panel < 0.1% * current range Voltage Distortion Factor: Superior to 0.5% Voltage Stability: Superior to 0.02% Sampling Current: 2 mA, 5 mA, 10 mA, 20 mA, 40 mA, 80 mA, 200 mA, 400 mA, 800 mA, 1.6 A, 4 A, 8 A (Peak Value) Sample Voltage: 20 mV, 40 mV, 100 mV, 200 mV, 400 mV, 800 mV, 2 V, 4 V, 8 V, 16 V, 40 V, 80 V, 160 V, 320 V, 800 V, 1600 V (Peak Value) | Conversion time: <= 2.5 μs (every channel) Resolution and Linearity: 16 Bits ± 1 LSB Voltage Range: ± 5 V ~ ± 10 V Sampling Clock: 6 μs ~ 10 ms Hardware Clock Internal Storage Capacity: 4 k Bytes Bus Structure: PCI |